Impact of low-dose CT scan in dual timepoint investigations: a phantom study

Abstract

Dual timepoint FDG takeup investigations have a potential for separating malignant lymph nodes from non-malignant in certain cases of suspected lung cancer. One hour seems to be the optimal time interval between the two scans (50-120 min). Many of the new PET scanners benefit from image fusion with a CT image and also use the CT for attenuation correction. In any practical hospital setting, 1 hour is too long to occupy the scanner bed and a second CT procedure thus becomes necessary. This study tries to validate to what extent the dose/quality of the second CT scan can be lowered, without compromising attenuation correction, lesion detection and quantification. Using a standard NEMA phantom with the GE Discovery PET/CT scanner, taken in and out between scan sessions, we have tried to find the minimal CT dose necessary for the second scan while still reaching tissue activity quantification within predetermined error limits. For a hot sphere to background activity concentration ratio of 1:5, the average uptake (normalised by the time corrected input activity concentration) in a sphere of 6 cm3 was found to be 0.90 ± 0.08 for the standard scan, yielding a dose of 5.5 mGy, and 0.90 ± 0.14 for a scan with lowest possible mAs product and lowest possible kV, yielding a dose of 0.65 mGy. With an insignificant increase in the uncertainty in the uptake measurement, we can get an order of magnitude reduction for the CT dose.

Info

Conference Paper, 2011